The Application of VMM3A Readout for Multi-Grid Neutron Detectors

A. Backis, F. Piscitelli, D. Pfeiffer, J. R. M. Annand, M. Aouane, K. G. Fissum, K. Livingston, G. Mauri, D. Raspino

公開日: 2025/10/2

Abstract

The T-REX neutron spectrometer at the European Spallation Source will use Multi-Grid Technology, which is a voxelised proportional counter relying on\mathrm{^{10}B_{4}C} coatings to detect the scattered neutrons. Measurements of the position dependence of pulse-height and relative detection efficiency of a Multi-Grid prototype of the T-REX spectrometer are presented for two different schemes of signal-processing electronics based on the VMM3A ASIC and CREMAT technology. These measurements, intended to test the suitability of VMM3A for readout of the T-REX Multi-Grid, are compared with Monte Carlo simulations based on the Garfield++ and Geant4 tool kits.