Xiao-Yong Liu, Peng Liu, Jie Li, Yu-Chen Zhang, Yuan-Bo Wang, Zhi-Peng Jia, Xiang Zhang, Xian-Qing Zhu, De-Quan Kong, Wen-Lan Song, Guo-Zhen Niu, Yu-Meng Yang, Pei-Jun Feng, Xiang-Pei Liu, Xing-Yang Cui, Ping Xu, Xiao Jiang, Juan Yin, Sheng-Kai Liao, Cheng-Zhi Peng, Han-Ning Dai, Yu-Ao Chen, Jian-Wei Pan

公開日: 2025/9/18

Abstract

Optical atomic clocks play a crucial role in fundamental physics, relativistic geodesy, and the future redefinition of the SI second. Standard operation relies on cyclic interrogation sequences, which alternate between atomic interrogation and dead time used for state preparation and readout. This approach introduces the Dick effect, where laser frequency noise aliases onto the atomic transition frequency. Although reducing laser noise improves clock stability, the Dick effect remains a key limitation. In this work, we demonstrate a zero-dead-time optical clock based on two interleaved ensembles of cold $^{87}\text{Sr}$ atoms. Our system significantly suppresses this noise and achieves a fractional frequency instability at the $10^{-19}$ level between 10,000 and 20,000 seconds over repeated measurements, with a best value of $2.9 \times 10^{-19}$ at $\tau = 20,000$ seconds. The estimated long-term stability based on the combined data of these measurements reaches $2.5 \times 10^{-19}$ at one day. These results represent a more than ninefold improvement over a conventional single-ensemble clock, highlighting its potential for next-generation timekeeping applications.